Test Code LAB798 Hepatitis A IgM (Anti-HAV, IgM), Serum
Performing Laboratory
Barnes-Jewish Hospital Laboratory
Specimen Requirements
Specimen Type: Serum
Container/Tube: Red/Grey separator or Gold separator tube
Specimen Volume: Full tube (Pediatric collection volume: 3.0 mL - minimum 2.0 mL)
Collection Information: Specimen with obvious microbial contamination is not acceptable.
Reference Values
Nonreactive (reported as reactive, nonreactive or equivocal). If result is equivocal, a new specimen should be submitted in 2 weeks for testing.
Interpretation depends on clinical setting.
Literature Reference:
Package insert: Roche Diagnostics, Elecsys Anti-HAV IgM 2022-11, V5.0
Methodology
Electrochemiluminescence Immunoassay (ECLIA)
Literature Reference:
Package insert: Roche Diagnostics, Elecsys Anti-HAV IgM 2022-11, V5.0
Additional Information
For BJH Laboratory Use Only
Alternate Tube Single test only(Not for Acute Hepatitis Panel):
1 pink-top (K2EDTA) tube
1 mint green-top (Lithium heparin) tube
1 green-top (Sodium heparin) tube
Separator gel tube is acceptable
Analyte Stability:
See Analyte Stability Chart under General Information for a complete list.
Additional Laboratory Information
Completed specimens are stored refrigerated for 5 days.
Test Classification and CPT Coding
86709
Specimen Transport Temperature
Ambient
Day(s) Test Set Up
Performed upon receipt
Turnaround Time:
STAT: not available
Routine: 2-4 hours after receipt in laboratory
LOINC Code Information
Name | LOINC Code |